LF Reader Plus (ATR-300),Get esay interpretation of COVID-19 Antigen & Antibody result.
Applications: |
√ Central laboratory of medical institutions |
√ Outpatient department |
√ Emergency department |
√ Medical services (such as community health centers) |
√Medical centers, etc. |
Features: |
● User-Friendly |
7"touch screen Principle |
Fast results in display |
Automatic print |
●Easy Operation |
Simple plug & play operation |
Less training required |
● Portable Design |
Smallin size (221*215*165 mm) |
Light in weight(<1.5 kg) |
● Multiple Test ltems |
Colloidal gold reagents |
Latex reagents |
Dry chemical reagents |
LF Reader Plus is designed to provide digital test results for in vitro diagnostic rapid test units, including Fertility Markers, Infectious Disease, Drug of Abuse, Tumor Markers and Cardiac Markers. It can be used with the dipstick, cassette and panel, which mainly used for the detection of colloidal gold reagents and/ or latex reagents. It can also be used for the detection of dry chemical reagents. The reader can be used for qualitative, semi-quantitative and quantitative analysis. For in vitro diagnostic use only.
LF Reader Plus uses a high-resolution camera to capture images of the test strip reflected on the object being tested, and the software algorithms automatically determine the test results and display them on the touch screen. The test results can be printed using the built-in thermal printer. The reader has the advantages of high precision and high speed, which can be used in the central laboratory of medical institutions, outpatient and emergency clinical departments or medical services (such as community health centers), medical centers, etc., and can also be used in scientific research laboratories.
For professional and Point of Care in vitro diagnostic use only.
Scope of Application
LF Reader Plus is intended for the use of reading in vitro diagnostic rapid tests manufactured by Citest that can examine different human samples with specific IVD tests. It is not designed to work with rapid tests manufactured by other manufacturers. Results obtained by using this equipment with test units manufactured by other manufacturers cannot be reliable.
External View

The power switch is on the rear side of the reader. Adapter connection port, LAN port and USB ports are on the left side for the purposes of connecting the reader with mains socket, computer and/or Laboratory Information System, scanner, mouse and so on.
Power adapter

1. Start the Reader
Connect power with Reader through the DC adapter and press the power switch to turn on the reader.
After initializing, fill in the username and password, the reader will have a self-testing. If the self-testing is passed, it will enter the main operation interface.
2. Locked
Within one day, when the Reader is not in use, it can be temporarily locked by click icon. When it is used again, fill in the username and password to enter the operation interface. We recommend that the instrument be turned off after using in one day
3.Shutdown
Press to turn off the reader. A prompt “Shut down the reader?” will be displayed on the screen, press “Yes” to turn off, press “No” to give up.
Note: Do not turn off the power switch immediately after the screen is turned off, about 1 minute later, the power switch can be turned off.
Warnings:
1) Failure to comply with the User Manual could lead to damage and inaccurate results.
2) Patient ID and Test Type cannot be edited once the test begins.
3) Do not open the Reader drawer during reading.
4) Do not reuse the used test product if a result has been reported.
5) Do not remove the USB drive during the software update.
Maintenance
Basic maintenance is to keep the outside of the Reader clean. External cleaning and maintenance method: Clean the outer surface of the reader with a damp cloth with no more than 70% alcohol. Do not use strong bleach ( ≥ 0.5% bleach solution), because oxidizing substances and solvents may damage the Reader’s parts and touchscreen. Do not clean any internal parts and inner surface.
Product features |
Parameters |
Format |
Dipstick, Cassette and Panel |
Reading Time |
<10 Seconds |
Pack |
1 Unite |
Storage Temperature |
-10~30℃ |